IBM Support

IJ45531: CALIBRATION WRONGLY ALLOWS AS FOUND STATUS TO BE CHANGED FROM FAIL TO PASS.

Subscribe to this APAR

By subscribing, you receive periodic emails alerting you to the status of the APAR, along with a link to the fix after it becomes available. You can track this item individually or track all items by product.

Notify me when this APAR changes.

Notify me when an APAR for this component changes.

 

APAR status

  • Closed as program error.

Error description

  • STEPS TO REPRODUCE:
    
    1. Go to Assets application.
    
    2. Create a new Calibration Asset record with any valid number,
    save and change Status to OPERATING.
    
    3. Go to Data Sheet Template application.
    
    4. Create a new Data Sheet Template with any valid number, 2
    Asset Functions and 1 Calibration Point each as the following:
    
    Asset Function: 1, Description: Asset Function 1, Calibration
    Point? selected, Type: DISCRETE, Repeatable? selected, No
    Adjustment Limit? selected
    Input Information: Tolerance on Input? selected, Min Field
    Resolution: 3, Units: Kg
    Tolerance 1: Single From: -0.300, To: 0.300, As: EU, Status when
    Exceeded: FAIL
    Calibration Point: 11, Description: Point 1-1, Nominal Input:
    Kg, Set Point: 19260.000 Kg
    
    Asset Function: 2, Asset Function 2, Calibration Point?
    selected, Type: ANALOG, Technician can Insert Points? selected,
    No Adjustment Limit? selected
    Input Information: Tolerance on Input? selected, Min Field
    Resolution: 1, Input Range: from 0 to 32100.0 Kg
    Output Information: Min Field Resolution: 1, Output Range: from
    0 to 32100.0 Kg
    Tolerance 1: Status when Exceeded: FAIL, Summed EU +: 0.3,
    Applied As: +/-
    Calibration Point: 21, Description: Point 2-1, Nominal Input:
    6420.0 Kg, Desired Output: 6420.0 Kg
    
    5. Save and approve the record.
    
    6. Go to WO Tracking application.
    
    7. Open a WO against the Calibration Asset created above.
    
    8. Go to Data Sheet tab and associate the Data Sheet created
    above.
    
    9. Save and approve WO.
    
    10. Still in the Data Sheet tab, select the Asset Function 2
    (ANALOG).
    
    11. In the Calibration Points section, enter the value 6420.0 in
    the As Found Input/Output and As Left Input/Output column
    fields.
    
    12. Save the record.
    
    Result:
    For Asset Function 2, both As Found and As Left Statuses change
    to PASS as expected.
    
    13. Still in the Data Sheet tab, select the Asset Function 1
    (DISCRETE).
    
    14. In the Calibration Points section, enter the value 100.000
    in the first As Found Set Point column field, just to simulate a
    fail. And enter 19260.000 in the other column fields just to
    pass.
    
    15. Save the record.
    
    Result:
    For Asset Function 1, As Found Status changes to FAIL as
    expected.
    
    16. Click on View Details icon of Asset Function 1.
    
    17. Scroll down until As Found Status section.
    
    18. Change the Status from FAIL to PASS.
    
    19. Save the record.
    
    RESULTS:
    The system allows the Status to be changed from FAIL to PASS,
    even without changing the entered value. This is the issue.
    
    EXPECTED RESULTS:
    The system should not allow the Status to be changed.
    
    REPORTED IN VERSION:
    IBM Maximo Asset Management 7.6.1.2
    IBM Maximo Calibration 7.6.1.0
    

Local fix

Problem summary

Problem conclusion

  • A fix for this APAR will be included in a future release.
    

Temporary fix

Comments

APAR Information

  • APAR number

    IJ45531

  • Reported component name

    MAXIMO CALIBRAT

  • Reported component ID

    5724R5700

  • Reported release

    761

  • Status

    CLOSED PER

  • PE

    NoPE

  • HIPER

    NoHIPER

  • Special Attention

    NoSpecatt / Xsystem

  • Submitted date

    2023-02-21

  • Closed date

    2023-03-02

  • Last modified date

    2023-03-07

  • APAR is sysrouted FROM one or more of the following:

  • APAR is sysrouted TO one or more of the following:

Fix information

  • Fixed component name

    MAXIMO CALIBRAT

  • Fixed component ID

    5724R5700

Applicable component levels

[{"Business Unit":{"code":"BU059","label":"IBM Software w\/o TPS"},"Product":{"code":"SSLKZS","label":"Maximo Calibration"},"Platform":[{"code":"PF025","label":"Platform Independent"}],"Version":"761","Line of Business":{"code":"LOB59","label":"Sustainability Software"}}]

Document Information

Modified date: